Description
TECHNICAL DATA
| Measured elements | carbon, sulfur |
| Samples | organic |
| Furnace alignment | horizontal (resistance furnace) |
| Sample carrier | ceramic boats |
| Field of application | agriculture, biology, chemistry / plastics, coal / power plant, construction materials, environment / recycling, medicine / pharmaceuticals |
| Furnace | resistance furnace (ceramic tube), adjustable up to 1550 °C (steps of 1 °C) |
| Detection method | solid state infrared absorption |
| Number of IR cells | 1 – 4 |
| Material of IR path | gold |
| Typical analysis time | 60 – 180 s |
| Chemicals required | magnesium perchlorate, sodium hydroxide |
| Gas required | oxygen 99.5 % pure (2 – 4 bar / 30 – 60 psi) |
| Power requirements | 230 V, 50/60 H, max. heating up current 20 A |
| Dimensions (W x H x D) | 58 x 52 x 75 cm |
| Weight | ~ 77 kg |
| Required equipment | PC, monitor, balance (resolution 0.0001g) |
| Optional accessories | TIC module |
CARBON / SULFUR ANALYZER ELEMENTRAC CS-RFUNCTIONAL PRINCIPLE
In the elemental analyzer CS-r the sample is burnt in an oxygen atmosphere at temperatures up to 1,550 °C. The furnace temperature can be freely selected in steps of 1 °C. The combustion gasses (CO2, H2O, SO2) coming from the furnace and pass through a dust filter. After the water vapor is chemically absorbed, the dried CO2 and SO2 gas is detected in the additional infrared cells. Depending on the configuration, it is possible to combine up to four (for C, S analyzers) infrared cells with different sensitivities.
Content may be subject to modifications or corrections